The C-MAC® Video Laryngoscope Is Superior to the Direct Laryngoscope for the Rescue of Failed First-Attempt Intubations in the Emergency Department.

TitleThe C-MAC® Video Laryngoscope Is Superior to the Direct Laryngoscope for the Rescue of Failed First-Attempt Intubations in the Emergency Department.
Publication TypeJournal Article
Year of Publication2014
AuthorsSakles JC, Mosier JM, Patanwala AE, Dicken JM, Kalin L, Javedani PP
JournalJ Emerg Med
Date Published2014 Dec 11
ISSN Number0736-4679
Abstract

OBJECTIVE: To compare the effectiveness of the C-MAC® video laryngoscope (CMAC) to the direct laryngoscope (DL) when used to rescue a failed first attempt intubation in the emergency department (ED).

METHODS: Data were prospectively collected on all patients intubated in an academic ED center over a five-year period from February 1, 2009 to January 31, 2014 when both the CMAC and the DL were available. Following each intubation the operator completed a continuous quality improvement (CQI) form documenting patient, operator and intubation characteristics. All orotracheal intubations attempted by emergency physicians (EPs) on adult patients with a failed first intubation attempt, and in which the CMAC or the DL was used for the second attempt, were included. The primary outcome was successful intubation on the second attempt using either the CMAC or the DL. A multivariate logistic regression analysis was performed to adjust for potential confounders.

RESULTS: During the five-year study period, there were 460 adult orotracheal intubation attempts by EPs which were not successful on the first attempt. In 398 (86.5%) of these cases the same operator performed the second attempt. The CMAC was utilized for the second attempt in 141 cases and was successful in 116 (82.3%; 95% CI 75.0%-88.2%) and the DL was utilized in 94 cases and was successful in 58 (61.7%; 95% CI 51.1%-71.5%). In a multivariate logistic regression analysis the CMAC was associated with an increased odds (adjusted OR 3.5; 95% CI 1.9-6.7) of a second attempt success compared to the DL.

CONCLUSIONS: After a failed first intubation attempt in the ED, regardless of the initial device used, the CMAC was more successful than the DL when used for the second attempt. This suggests that the CMAC is the preferred rescue device after an initial intubation attempt in the ED fails.

DOI10.1016/j.jemermed.2014.10.007
Alternate JournalJ Emerg Med
PubMed ID25498851
Faculty Reference: 
Jarrod Mosier, MD
John C. Sakles, MD, FACEP